Digital Pressure Meter with Directly Connected Sensor

Digital Pressure MeterHigh Accurate Pressure Intelligent Digital Display

VSW2Hseries

500kPa~50MPa
High Accurate Pressure Intelligent Digital Display
* Actual products may differ from photo
  depending on specifications.

High accuracy: ±0.25%F.S. ±1 digit or below (Service temp.: 0 to 50°C, Overall accuracy including temperature characteristics)
"Max./min memory" function included, which allows abnormal pressures detected so far to be checked during operation.
Color LCD adopted with display colors "red and green" intuitively understandable for users.
Setting output can be selected between higher/lower limits and NPN/PNP (Factory-set at NPN/PNP.)

Measured medium VTsensor: Water, oil, gases or other media that will not corrode SUS630 and SUS430.
Hsensor: Water, oil, gases or other media that will not corrode Hastelloy C22 and SUS316L.
Pressure type Positive and compound pressures can be measured.
Pressure range 500kPa~50MPa

* EAC Mark is an option item: EAC Mark option includes TECHNICAL RULES OF THE CUSTOMS UNION (TR-CU) Declaration-of-conformity document only, or includes both of TECHNICAL RULES OF THE CUSTOMS UNION (TR-CU) Declaration-of-conformity document and Metrology Pattern Approval certificate (MPA).
* The products in the photo are standard specification models. Please contact our sales office for further details.

Specification

Display -1999~2000(*1) 7 segments 4 digit LCD(Character height:9mm)
Display Color Red or Green (Arbitrary setting)
Rated Capacity 500kPa, 1, 2 , 5 , 10 , 20 , 35 , 50MPa
Calibration Range 0~(500kPa, 1, 2 , 3.5 , 5 , 10 , 20 , 35 , 50)MPa
Accuracy ±0.25%F.S.±1digit or less
[Non Linearity, hysteresis, repeatability, Temperature characteristics of 0 to 50°C(25°C standard) included]
Output Channels 2 points of Open collector output (NO/NC selectable)
Setting output Maximum rating DC30V 20mA Residual voltage: 0.85V (NPN output)
1 points of Open collector output + 4~20mA output
(Safe load resistance: 400Ω or less)
Use Temperature Range Temperature: 0~50°C(Temperature of medium to be measured: -10~70°C)
Safe Overload Rating 150%R.C.(*2)
IP Rating Equivalent to IP65(Dedicated cable connection state)
Temperature Effect Included in Accuracy above
Vibration-Proof 98m/s²(10G)[Compliant with IEC68-2-6]
Connection Screw Rc1/4[Female screw]: (Option)R1/4[Male screw]
RoHS Europe RoHS directive conformity product
Connector Model number XS2M-D423 made in Omron
Measurement Medium Water, oil, gases or other media that will not corrode SUS630 and SUS430.
Power Source Voltage DC24V±10%
Consumption current Approx. 30mA(Approx. 50mA for models with analog output)
Functions Display oscillating mechanism(±120°)/ 2nd display function / Display update rate
Zero shift / Zero suppress / Simulation / Damping
FIX zero / Number of moving average / Peak hold / Key protection
Maximum and Minimum memory
Pressure Port Material VTsensor: SUS630
Hsensor: Hastelloy C22
Connection Screw Material VTsensor: SUS430
Hsensor: SUS316L
Weight Main body: Approx. 230g Dedicated cable 2m: Approx. 60g
Accessories Operation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy)
Options (*3) Wall hangings fixation bracket [VSW2-BR]
  • (*1) Depends on the factory setting. Values beyond the calibration range cannot be displayed.
  • (*2) R.C.: Rated Capacity
  • (*3) Contact our sales office for details.

Model Selection

Model Selection
Selection Check Model Specification
1.Model VSW2H High Accurate Pressure Intelligent Digital Display
2.Output   02 2 points of Open collector output (Factory setting: NPN)
  31 1 points of Open collector output (Factory setting: NPN)
+4~20mA output (Scaling setting possible as desired.)
3.Power 4 DC24V±10%
(Consumption current: Approx. 30mA *Analog output: Approx. 50mA)
4.Calibration   500kPa (0~500kPa) Sensor M designation Rated capacity 500KPG designation
  1000kPa (0~1000kPa) Rated capacity 001MPG designation
  1.000MPa (0~1.000MPa)
  2000kPa (0~2000kPa) Rated capacity 002MPG designation
  2.000MPa (0~2.000MPa)
  3.50MPa (0~3.50MPa) Rated capacity 3R5MPG designation
  3.500MPa (0~3.500MPa)
  5.00MPa (0~5.00MPa) Sensor VT designation Rated capacity 005MPG designation
  5.000MPa (0~5.000MPa)
  10.0MPa (0~10.0MPa) Rated capacity 010MPG designation
  10.00MPa (0~10.00MPa)
  20.0MPa (0~20.0MPa) Rated capacity 020MPG designation
  20.00MPa (0~20.00MPa)
  35.0MPa (0~35.0MPa) Rated capacity 035MPG designation
  35.00MPa (0~35.00MPa)
  50.0MPa (0~50.0MPa) Rated capacity 050MPG designation
  50.00MPa (0~50.00MPa)
5.Cable Connection   WN Cable not included. (If prepared by the customer),
Front connector type
Main body use connector: Panel mounting connector plug made in Omron
(Model number)XS2M-D423
  WS Waterproof connector (Front: Straight type) + Cable 2m
  WR Waterproof connector (Front: Angle type) + Cable 2m
  RN Cable not included. (If prepared by the customer),
Back connector type
Main body use connector: Panel mounting connector plug made in Omron
(Model number)XS2M-D423
  RS Waterproof connector (Back: Straight type) + Cable 2m
  RR Waterproof connector (Back: Angle type) + Cable 2m
6.Sensor Type   H Semiconductor thin-film type sensor
  VT Thin-film sensor
7.Rated Capacity(R.C.)   500KPG 500kPa
  001MPG 1000kPa
  002MPG 2000kPa
  3R5MPG 3500kPa
  005MPG 5MPa
  010MPG 10MPa
  020MPG 20MPa
  035MPG 35MPa
  050MPG 50MPa
8.Connection screw   C2 Rc1/4
  R2 R1/4

Learn more about external dimensions.


.